Unimodule "METD" 100% Exhausted Vertical Flow Workstation

Introduction

Designed for operations involving toxic or noxious fumes that can be potentially harmful to workers; a rear plenum and workstation exhaust slots remove exhaust fumes from within the cabinet. The open front features vertical airflow for personnel protection; the HEPA filter ensures product protection.

Features

  • ISO Class 5 (Class 100) clean air work zone.
  • 100% exhausted air used for application with toxic or noxious fumes.
  • Provides personnel and product protection.
  • Sizes: 4', 6' and 8'.

Other Related Products



† Proper unit operation requires an exhaust system (provided and installed by others) capacity equal to 125% of the face velocity and air supply volume. Makeup air must compensate for that exhausted with final exhaust balancing achieved by a customer provided and installed damper in the external exhaust duct.

* As defined by the latest ISO 14644 and IEST Recommended Practices, where particle count does not exceed 3520 particles, 0.5 µm or larger, per cubic meter of air.

For additional information, please contact your local representative.